Pediatrics and Adults

  One Device for All Patients

It  is  suitable  for  all  pediatric,  adult  male  and  female  patients.

While  lowering  the  tidal  volume  down  to  2  cc,  it  can  increase

breath  rate  up  to  150  breath/minute.

Calculates  patient  elastance  and  compliance  values  fast,

correctly  and  precisely.

Automatically  calibrates  at  the  start.

Due  to  the  automatic  tube  compensation,  it  provides

airway  resistance  and  patient  circuit  volume  compensation.

It  provides  80%  leakage  compensation  by  modelling  patient

circuit  leakages.

Enables  Inspiration  Hold  and  Expiration  Hold  functions  that  are

adjustable  between  1-60  sec.

Doctor's Clinic
IMG_3112.jpg
Advanced Adaptive Control

Device uses advanced adaptive control algorithms.

Responds in milliseconds

Device  has  modes  for  Nasal  Cpap  and  High  Flow  Oxygen  Therapy.

 

Device  has  integrated  nebulizer  system.

Device operates in conventional and innovative modes.

Pressure  Controlled Modes:

 

P-CMV

V-CMV

PRVC-CMV

Volume  Controlled Modes:

 

P-SIMV

V-SIMV

PRVC-SIMV

Spontaneous  and  Smart Modes:

PSV

VSV

CPAP

Device Interface and Software Features

  15-inch  Full  Touch  Screen

Device displays system information on 15-inch high definition colorful screen and satisfies usage with sensitive touch ability. Quick access on the screen for parameters like Fio2, PEEP, Breath and Respiration. Via its wide screen,device shows real-time patient information and allows you to know any alarm situation while you are distant. Device interface has a user-friendly design. Provides an easy use opposed to intensive care unit mechanical ventilator devices that are known as complicated.

Hand%20_edited.png
PATIENT
COMPATIBILITY
AND PERFORMANCE

Child, Pediatric

Breath Rate: 1 - 150 / Min

 

T Inspiration: 0.1 – 10 Secs

 

Tidal Volume: 0.02 – 0.6 L

 

Flow: 1 – 60 L/Min

 

Trigger Sensitivity:

0.1 – 20 L/Min

0.1 – 20 mbar

 

Leakage and Tube Compensation

 

Invasive and Non-Invasive Ventilation

Adults

Breath Rate: 1 - 100 / Min T

 

Inspiration: 0.1 – 10 Secs Tidal

 

Volume: 0.1 – 3 L Flow: 1 – 120

 

L/Min Trigger Sensitivity: 0.1 – 20

 

L/Min 0.1 – 20 mbar Leakage and

 

Tube Compensation Invasive and

Non-Invasive Ventilation

Clínica

Ventilation Modes

P-CMV

V-CMV

PRVC-CMV

P-SIMV

V-SIMV

PRVC-SIMV

PSV

VSV

CPAP

Pressure Continuous Mandatory Ventilation

Volume Continuous Mandatory Ventilation

Pressure Regulated Volume Control

Pressure Controlled Synchronized Mandatory Ventilation

Volume Controlled Synchronized Mandatory Ventilation

Pressure Regulated Volume Control

Pressure Support Ventilation

Volume Support Ventilation

Continuous Positive Airway Pressure

Operational Properties

Inspiration Pressure

Inspiration Time

PEEP

Breath Rate

Tidal Volume

Flow Rate

O2 Concentration

Spontaneous Pressure Support

I/E Ratio

2-100 mBar

0.1-10 secs

1-50 mBar

(p): 1-150/min

(a): 1-100/min

 

(p): 20-600 mL

(a): 100-3000 mL

 

(p): 1-60 l/min

(a): 1-120 l/min

 

21-100%

0-100 mBar

1:10(x60*)-10:1

 

(p): Pediatrics, (a): Adults

Detailed Properties

Apnea Time

Apnea Ventilation

Apnea Mode

Pressure Trigger

Inspiration Stop

Tramp

Automatic Tube Compensation

Automatic Leakage Compensation

 

Trigger Compensation

Inspiration Pause

Expiration Pause

Integrated Nebulizer

O2 Support

5-60 secs, Double sided

Apnea Ventilation

P-CMV, V-CMV

0.1-20 l/min

0.1-20 mBar

0-80%

0.1-5 secs

0-80%

0-80%

 

On-Off

 

1-60 secs

1-60 secs

1-20 L/Min

 

2 Mins

Clínica

Monitored Parameters

P Peak

P Peep

P Plato

P Average

F Inspiration

F Expiration

MVe

SpnMVe

SpnMVe/MVe

V Tidal

FiO2

Breath Rate

Spontain Breath Rate

T İnspiration

T Exspiration

I/E Ratio

R Airway

C Static

C Dynamic

Elestance

RC Constant

WOB

WOB/L

V Exspiration

V Ads

AutoPeep

P0.1

RSBI

PTP

FTP

P NIF

MVsp%

Leakage Ratio

Leakage Volume

Peak Inspiratory Pressure Measurement

Peep Pressure Measurement

Inspiratory Plato Pressure Measurement

Average Pressure Measurement

Inspiratory Flow Rate

Expiratory Flow Rate

Minute Volume Measurement

Spontaneous Minute Volume Measurement

Spontaneous Minute Volume/ Minute Volume Ratio Tidal Volume

Oxygen Ratio

Minute Breath Rate

Spontaneous Minute Breath Rate

Inspiratory Time

Expiratory Time

Inspiratory Time/ Expiratory Time Ratio

Airway Resistance Measurement

Static Compliance Measurement

Dynamic Compliance Measurement

Elastance Measurement

RC Time Constant Measurement

Work of Breathing (Energy consumed during Inspiration)

Work of Breathing/Volume Ratio

Expiration Tidal Volume

Anatomical Dead Volume Measurement

End of Respiration Pressed Air Pressure

100msec Occlusion Pressure Measurement

Rapid Shallow Breathing Index

Negative Pressure x Negative Pressure Time

Negative Flow x Negative Flow Time

Negative Inspiratory Pressure Force

Spontaneous and Mandatory Minute Volume Ratio

Leakage Volume Ratio

Soluk End of Breath Cycle Leakage Volume

Pressure Source Properties

O2 Pressure

Air Pressure

2.5-7 Bar Central System/Tube 2.5/7 Bar Central System/Tube

Automated Replacement and Alarm Display when the Source Finishes.

Compatible with Medical Compressor/Regulator

Graphical Properties

Pressure-Time Graphics Volume-Time Graphics Flow-Time Graphics Pressure-Volume, Pressure-Flow, Flow-Volume Loops

Size & Weight

Height

Depth

Width

Weight

Screen Movement

Screen

142cm (4.65 ft)

53cm (1.74 ft)

46cm (1.50 ft)

48kg (106 lb)

Stable

15-inch Full Touch

Digital Interfaces

2 USB, 1 COM, 1 Ethernet

Alarm Properties

Voice and Video Alarms and Alarm Configuration 2 minutes Alarm Mute

Inspiratory Pressure

Tidal Volume

Respiration Rate

Minute Volume

I/E Ratio

FI02

Apnea Time

Leakage

Lower Limit / Upper Limit

Lower Limit / Upper Limit

Lower Limit / Upper Limit

Lower Limit / Upper Limit

Lower Limit / Upper Limit

Lower Limit / Upper Limit

Upper Limit

Upper Limit

Electrical Properties

Battery Duration

Grid Voltage

,Power Consumption

2 Hours

+8 Hours, Optional

180-264VAC,

47-63Hz 112W

Patient Recording and Logging

• 20+ Ventilation Value Trend Records for the last 3 days and Graphical Display

• 2000 Records Alarm and Ventilation Settings Information Storage